# FDA PCCP for AI/ML medical devices: a builder's guide

> **Informational only — not legal advice.** Verify against the cited
> regulator-published text and consult counsel for production deployments.
> See `AI-DISCLOSURE.md` in this package.

If you're building an AI/ML-enabled medical device or device software
function — a clinical decision-support tool that gets cleared by FDA,
an imaging algorithm in a 510(k)-cleared scanner, an AI radiology
triage system, an AI-driven continuous glucose monitor, or any other
software-as-a-medical-device (SaMD) that uses machine learning — the
**FDA Predetermined Change Control Plan (PCCP) framework** is the
specific federal regulatory vehicle that lets you iterate on the
model after authorization without filing a new submission for every
update. This guide covers what § 515C of the FD&C Act actually
requires, what a PCCP looks like in production, the labeling and
public-summary disclosure obligations that come with it, how it
stacks with HHS Section 1557 and state-level rules, and what
governance any AI/ML device team needs in place before submission.

## What FDA PCCP actually is

The Federal Food, Drug, and Cosmetic Act § 515C (21 U.S.C. § 360e-4)
was added by **Section 3308 of the Food and Drug Omnibus Reform Act
of 2022** (FDORA, P.L. 117-328). It authorizes FDA to clear or
approve a Predetermined Change Control Plan as part of an AI/ML
device's marketing submission — meaning: the manufacturer pre-
specifies the kinds of modifications it intends to make to the AI/ML
algorithm post-authorization, the methods it will use to validate
those modifications, and the assessment of their impact. Once FDA
authorizes the PCCP, the manufacturer can implement modifications
that conform to the plan without a new marketing submission.

On December 4, 2024, FDA issued the [final guidance](https://www.fda.gov/regulatory-information/search-fda-guidance-documents/predetermined-change-control-plans-artificial-intelligence-enabled-device-software-functions),
*"Predetermined Change Control Plans for Artificial Intelligence-
Enabled Device Software Functions."* The final guidance applies to
all medical devices regardless of pathway (510(k), De Novo, PMA)
and supersedes the April 2023 draft. It is the authoritative
reference for what a PCCP must contain, how to validate
modifications, and how to disclose the AI/ML nature of the device
to clinicians and patients.

The framework solves a real problem. Before § 515C, any change to
the algorithm of a cleared or authorized AI/ML device that affected
the device's safety or effectiveness typically required a new
510(k), De Novo, or PMA submission. Iterative model improvement
became impractical: every meaningful retrain triggered a new
regulatory cycle. PCCP lets manufacturers pre-authorize a bounded
set of modifications (and the validation methods for each) so
iteration can happen within the bounds the agency has reviewed.

## What a PCCP must contain

Per the final guidance, every PCCP comprises three components:

### 1. Description of Modifications

A specific list of the modifications the manufacturer intends to
make to the AI-enabled device software function under the PCCP.
Each modification must be:

- **Specific.** "We may improve the algorithm" is not a modification
  description. "We may retrain the model on additional pediatric
  data drawn from the same patient population, with retrained
  weights deployed only after the validation in the Modification
  Protocol shows non-inferior sensitivity and specificity at the
  authorized device's operating point" is.
- **Bounded.** The set of permissible modifications is finite. The
  PCCP must enumerate them; modifications outside the enumeration
  require a new marketing submission.
- **Predictable in impact.** The Description of Modifications
  pairs with the Impact Assessment to show that the predicted impact
  is positive or neutral, and that risks have been characterized.

Common modification categories:
- Retraining on additional data (with bounded data-distribution
  assumptions).
- Updates to feature engineering or input preprocessing.
- Threshold adjustments at the operating point.
- Performance improvements on specific subgroups.
- Compatibility updates for new sensor inputs.

### 2. Modification Protocol

Methods to develop, validate, and implement the planned modifications.
The Modification Protocol is the testable specification of how the
manufacturer will know whether a proposed modification meets the
required performance bar. It must include:

- **Data management.** What data will be used for retraining; how
  it's sourced; how data quality is maintained; how patient
  populations are represented.
- **Retraining methodology.** The algorithmic procedure used to
  produce a candidate modified model.
- **Performance evaluation.** The metrics the modified model must
  meet — typically including sensitivity, specificity, AUC, and
  fairness across demographic subgroups — and the operating points.
- **Update procedures.** How the modification is deployed to the
  device, including version control, rollback, and clinician
  notification.

The Modification Protocol is the most consequential part of a PCCP.
A weak Modification Protocol can result in FDA limiting the PCCP's
scope or refusing authorization.

### 3. Impact Assessment

Evaluation of the benefits and risks of each anticipated modification,
including:

- **Benefit characterization.** What the modification is intended to
  improve and how it will be measured.
- **Risk characterization.** Foreseeable risks the modification
  introduces, and the controls that will detect or mitigate them.
- **Cumulative-impact analysis.** Where multiple modifications could
  compound, the assessment must consider their combined effect.
- **Comparison against the authorized baseline.** Each modification
  must perform at least as well as the originally authorized device
  on the metrics that drove the original authorization.

## Labeling and public-disclosure obligations

PCCP doesn't change the underlying labeling regime under 21 CFR
Part 801; it adds specific disclosure expectations on top.

The device labeling (which includes the user manual, the
manufacturer's product page, and FDA's public-facing 510(k) Summary,
De Novo Decision Summary, or PMA Approval Order) must:

1. **Disclose the AI/ML nature** of the device. State that the
   device is an AI-enabled device software function and identify
   the regulatory pathway and submission number.
2. **Summarize the PCCP** where one is authorized. State the bounds
   of the modifications that may be implemented without a new
   submission.
3. **Inform clinicians** that the device may be modified within the
   PCCP without further FDA review.
4. **Provide a current device summary** that reflects the current
   model version, the validation data for that version, and the
   cumulative record of PCCP-conforming modifications implemented to
   date.

A public-facing device-summary page, updated each time a PCCP-
conforming modification is implemented, is the de facto best practice
emerging from the December 2024 final guidance. FDA's own public-
facing pages (510(k) Summary, etc.) reflect the original
authorization; the manufacturer page is where current model state
lives.

Plain-language template that satisfies the labeling requirements:

> *"This device incorporates an artificial intelligence or machine-
> learning algorithm. The device has been authorized for marketing
> by the U.S. Food and Drug Administration under [510(k) / De Novo
> / PMA number]. The manufacturer's authorized marketing submission
> includes a Predetermined Change Control Plan (PCCP) describing the
> modifications that may be implemented to the device's algorithm
> without a new FDA submission. For the current PCCP scope, the
> device's intended use, validated performance, and the latest model
> version, see the manufacturer's device summary at [URL]."*

## How PCCP applies across pathways

The final guidance applies to all device-pathway pathways, but the
mechanics differ slightly:

| Pathway | When PCCP fits | Common AI/ML device classes |
|---|---|---|
| **510(k)** (substantial equivalence) | PCCP filed alongside the 510(k) submission; FDA reviews and authorizes within the 510(k) timeframe. | Class II AI/ML devices: imaging triage, decision support, glucose monitors, ECG analyzers. |
| **De Novo** (low-to-moderate-risk novel device) | PCCP filed in the De Novo request; authorized as part of the request. | Novel AI/ML diagnostics with no predicate device. |
| **PMA** (premarket approval, Class III) | PCCP filed in the PMA module; supplemental approval. | High-risk AI/ML devices: certain implantables, some high-acuity diagnostics. |

The 510(k) pathway is by far the most common for AI/ML devices —
about 95% of FDA-authorized AI/ML medical devices are 510(k)-cleared.

## How PCCP stacks with HHS Section 1557

Section 1557's Patient Care Decision Support Tool (PCDST)
nondiscrimination obligations (45 CFR § 92.210, effective 2025-05-01)
operate at the **deployer** level — the covered entity that uses the
device. PCCP operates at the **manufacturer** level — the entity
that builds and authorizes the device.

Both apply to the same AI/ML medical device:

- **Manufacturer obligations** under FDA: PCCP-bounded modifications,
  labeling disclosure, post-implementation transparency, ongoing
  performance monitoring under 21 CFR Part 803 (medical device
  reporting).
- **Deployer obligations** under HHS Section 1557: PCDST inventory,
  mitigation of discrimination risk, designated Civil Rights
  Coordinator coverage, patient-facing notice where applicable.

A hospital using an FDA-cleared AI radiology triage tool: the
manufacturer's PCCP governs how the tool is updated; the hospital's
Section 1557 PCDST process governs whether and how the tool is used,
and how the hospital monitors for discriminatory output. Both
obligations apply. See the [HHS Section 1557 builder's guide](/guides/hhs-section-1557-pcdst-builder-guide/)
for the deployer side.

## How PCCP stacks with state laws

| State rule | How it stacks |
|---|---|
| **California SB 1120 (Physicians Make Decisions Act)** | Effective 2025-01-01. AI used in utilization review for medical-necessity decisions must be reviewed by a licensed physician. Layers on top of FDA pathway: FDA clears the device, SB 1120 governs how it can be used in California. |
| **NYDFS October 2024 cybersecurity / AI guidance** | Applies to NYDFS-licensed entities. AI tool risks must be addressed in cybersecurity programs. AI/ML medical devices held by NY-licensed insurers fall in scope. |
| **State medical-board AI rules** (TX, several others) | Govern how clinicians may use AI in scope of practice. Layer on top of the manufacturer-level FDA framework. |

The right rule for production deployment is the strictest applicable
overlay, not FDA alone.

## How the public-facing device summary should evolve

The December 2024 final guidance treats post-implementation
transparency as integral to PCCP compliance. The public-facing
device summary on the manufacturer's site is the practical surface.
What it should contain:

- **Current model version.** A version identifier the clinician can
  cross-reference against the device labeling.
- **Date of last modification.** When the most recent PCCP-conforming
  change was implemented.
- **Validation data for the current version.** Performance metrics
  on the validation set, including subgroup performance where the
  device is intended for diverse patient populations.
- **PCCP scope.** The bounds of authorized modifications, summarized
  for non-regulator readers.
- **Cumulative modification log.** A chronological list of PCCP-
  conforming modifications implemented since authorization.
- **Contact for questions.** A path for clinicians and patients to
  reach the manufacturer about the AI/ML nature of the device.

A device summary that omits these elements is not yet aligned with
the final guidance's expectations. Expect FDA to lean on this in
post-market surveillance.

## Common compliance failure patterns

- **Modifications outside the authorized PCCP.** A retraining run
  that uses a data source not covered in the Description of
  Modifications. Even if the resulting model is "better," it
  requires a new marketing submission.
- **Modification Protocol that doesn't enforce its own metrics.** A
  PCCP whose Modification Protocol describes validation but doesn't
  state explicit pass/fail thresholds. FDA may treat post-
  authorization changes as outside the PCCP's scope.
- **No public-facing device summary.** Device labeling references a
  PCCP but the manufacturer doesn't provide an updatable public
  summary; clinicians can't tell what model version is currently
  deployed.
- **Section 1557 deployer obligations treated as the manufacturer's
  responsibility.** The covered entity (hospital, FQHC, etc.) is
  responsible for its own PCDST inventory and mitigation —
  the manufacturer's FDA labeling does not satisfy the deployer's
  HHS Section 1557 obligations.
- **Cumulative-impact analysis missing.** PCCP allows multiple
  modifications. Without a cumulative-impact assessment, drift over
  many modifications can leave the device performing meaningfully
  differently from the originally authorized baseline.
- **Fairness / subgroup performance not in the Modification
  Protocol.** A PCCP whose Modification Protocol only checks aggregate
  performance metrics misses subgroup-level performance changes.
  These can trigger Section 1557 disparate-impact concerns at the
  deployer level — and create FDA postmarket safety issues.

## How plainstamp helps

`plainstamp` ships a `us-fda-pccp-aiml-device-software-2024` rule
that returns the labeling-disclosure checklist, plain-language and
formal-language device-labeling templates, citation back to FD&C Act
§ 515C and the December 2024 FDA final guidance, and a
`last_verified` date. Lookup:

```bash
npx plainstamp lookup --jurisdiction us \
                      --channel about-page \
                      --use-case healthcare
```

For California-operating manufacturers, layer SB 1120 on top:

```bash
npx plainstamp lookup --jurisdiction us-ca \
                      --channel about-page \
                      --use-case healthcare
```

## The minimum viable compliance posture

If your AI/ML medical device is starting from zero on PCCP / labeling
compliance, ship these six artifacts in order:

1. **Authorized PCCP** in your marketing submission. Description of
   Modifications, Modification Protocol with explicit pass/fail
   thresholds, Impact Assessment with cumulative-impact analysis.
2. **Device labeling** that discloses the AI/ML nature, summarizes
   the PCCP, and points to the public-facing device summary URL.
3. **Public-facing device summary page** with current model version,
   date of last modification, validation data for the current
   version, PCCP scope, cumulative modification log, contact path.
4. **Modification implementation runbook.** A documented procedure
   for going from "candidate modification" to "deployed PCCP-
   conforming modification": validation against the Modification
   Protocol, version-control update, labeling/summary update,
   clinician notification, audit-trail entry.
5. **Subgroup performance monitoring.** Ongoing monitoring that
   detects performance drift overall AND across protected-class
   subgroups, with thresholds that escalate to a new marketing
   submission if exceeded.
6. **Coordination path with deployers.** A documented contact and
   escalation channel for hospital / FQHC / insurer customers
   who need to satisfy their Section 1557 PCDST obligations.

Then layer the higher-fidelity work — postmarket surveillance under
21 CFR Part 803, risk-class-specific quality-system requirements
under 21 CFR Part 820, sector-specific overlays — onto the higher-
risk modification categories first.

## Source-of-truth links

- **FDA Final Guidance — PCCP for AI-Enabled Device Software Functions (December 2024)** ([fda.gov](https://www.fda.gov/regulatory-information/search-fda-guidance-documents/predetermined-change-control-plans-artificial-intelligence-enabled-device-software-functions))
- **FD&C Act § 515C, 21 U.S.C. § 360e-4** ([uscode.house.gov](https://uscode.house.gov/view.xhtml?req=granuleid:USC-prelim-title21-section360e-4&num=0&edition=prelim))
- **FDA Modernization Act of 2022 / FDORA (P.L. 117-328 Division FF Title III)** ([congress.gov](https://www.congress.gov/bill/117th-congress/house-bill/2617))
- **21 CFR Part 801 (Device Labeling)** ([ecfr.gov](https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-801))
- **FDA AI/ML-enabled medical device list** ([fda.gov](https://www.fda.gov/medical-devices/software-medical-device-samd/artificial-intelligence-and-machine-learning-aiml-enabled-medical-devices))